Sterilization Warning • Sterilize by autoclave after each patient use. Sterilization is required before first use. • Wipe off handpiece with damp cloth. • Place the handpiece in a sterilization pouch and seal. Bag handpiece individually. • Steam autoclave for 15 minutes at 275° F (135° C) in an autoclave. • Allow a minimum of 20 minutes for the drying time. Caution • Do not use ultrasonic bath, boiling water or chemicals to clean the handpiece. • Do not autoclave handpieces together with other instruments. • Do not use dry heat sterilization. |
![]() |