Repeated treatment cycles including manual washing and sterilization have minimal effects on Newtron LED handpiece scaled handpieces, LED ring and optical guide.
Repeated treatment cycled that include ultrasonic cleaning, manual washing and sterilizations have minimal effects on Scaler Handpiece Nosepieces.
End of service life is normally determined by heat and damage due to use.
Important
• Throughout the procedure, wipe blood and/or debris from the Handpiece to prevent it drying on the surfaces.
• In case of external irrigation tubing or configuration of the device with tank, after each use, rinse the irrigation tubing, Handpiece and tip for 1 minute with an enzymatic or alkaline cleaning solution to prevent debris and contamination from drying on the device. Rinse with distilled or purified water for 1 minute to remove the enzymatic or alkaline cleaner. Flush the irrigation tubing to remove liquid from the inside.
• In the case of irrigation coming from the dental practice water network, clean and rinse the cord, the Handpiece and the tip 1 minute after each use.
• Soiled devices should be separated from non-contaminated devices to avoid contamination of personnel or surroundings.
• The Handpiece must be covered with a cloth dampened with purified water to prevent blood and/or debris from drying on the surface.